Home
preclinical-testing-for-medical-devices-under-mdr

Preclinical Testing for Medical Devices under MDR

Preclinical Testing for Medical Devices under MDR: A Comprehensive Overview

The Medical Device Regulation (MDR) has been in effect since May 2021, bringing significant changes to the regulatory landscape for medical devices across Europe. As part of these new regulations, preclinical testing plays a crucial role in ensuring that medical devices meet the required safety and performance standards. In this article, we will delve into the world of preclinical testing under MDR, exploring the requirements, best practices, and key considerations for manufacturers.

Understanding Preclinical Testing under MDR

Preclinical testing is an essential step in the development process of medical devices, aimed at assessing their safety and efficacy before human clinical trials. Under MDR, preclinical testing is mandatory for all medical devices, with the exception of those that are exempt due to low-risk classification (Class I). The primary objectives of preclinical testing include:

  • Evaluating the devices biocompatibility and toxicology

  • Assessing its mechanical performance and durability

  • Investigating any potential interactions or adverse effects on human tissues


  • Manufacturers must ensure that their preclinical testing is conducted in accordance with established standards, guidelines, and regulatory requirements. This includes adhering to ISO 10993-1 (Biological evaluation of medical devices - Part 1: Evaluation and testing), which provides a framework for assessing the biocompatibility of medical devices.

    Key Considerations for Preclinical Testing

    Several key considerations are crucial when conducting preclinical testing under MDR:

  • Study Design: The study design should be robust, reliable, and relevant to the intended use of the device. This includes selecting the appropriate animal models, test conditions, and endpoints.

  • Test Methods: Manufacturers must choose suitable test methods that align with the ISO 10993 series standards. These methods include cytotoxicity testing (ISO 10993-5), sensitization testing (ISO 10993-10), and genotoxicity testing (ISO 10993-3).

  • Data Analysis: Results should be analyzed in accordance with statistical guidelines, ensuring that conclusions are based on valid and reliable data.


  • Detailed Information on Preclinical Testing Requirements

    Here is a more detailed explanation of the preclinical testing requirements under MDR:

    Biocompatibility Testing

    Biocompatibility testing involves evaluating the potential interactions between the device and human tissues. This includes assessing cytotoxicity, sensitization, and genotoxicity.

  • Cytotoxicity Testing: This test evaluates the devices potential to induce cell death or damage.

  • Cell culture methods (ISO 10993-5)

    In vivo testing using animals

  • Sensitization Testing: This test assesses the devices ability to cause an immune response, leading to skin sensitization.

  • Guinea pig maximization test (GPMT)

    Mouse local lymph node assay (LLNA)

  • Genotoxicity Testing: This test evaluates the devices potential to induce genetic damage or mutations.

  • In vitro testing using bacterial and mammalian cell lines

    In vivo testing using animals

    Mechanical Performance and Durability

    In addition to biocompatibility, preclinical testing also involves assessing the mechanical performance and durability of medical devices.

  • Tensile Testing: This test evaluates the devices tensile strength, elongation at break, and other mechanical properties.

  • Fatigue Testing: This test assesses the devices resistance to cyclic loading or fatigue.

  • Impact Testing: This test evaluates the devices ability to withstand impacts without failing.


  • QA Section

    This section provides additional information on preclinical testing under MDR:

    Q: What are the main differences between MDD and MDR?

    A: The Medical Device Directive (MDD) has been replaced by the Medical Device Regulation (MDR). While both regulations have similar objectives, MDR introduces several changes, including increased transparency and accountability.

    Q: How do I choose the right animal model for my preclinical testing?

    A: When selecting an animal model, manufacturers should consider factors such as species specificity, sample size, and experimental design. Consultation with experts in the field is recommended to ensure that the chosen model aligns with the devices intended use.

    Q: What are some common challenges associated with conducting preclinical testing under MDR?

    A: Challenges may arise from issues such as:
  • Insufficient resources or expertise

  • Limited understanding of regulatory requirements

  • Difficulty in selecting suitable animal models or test methods


  • Q: Can I perform preclinical testing on my own, or is it necessary to outsource to a third-party laboratory?

    A: While manufacturers can conduct preclinical testing in-house, outsourcing to a third-party laboratory may be more cost-effective and time-efficient. However, ensure that the chosen laboratory has experience with MDR requirements and adheres to established standards.

    Q: What happens if my device fails to meet preclinical testing requirements under MDR?

    A: If a device fails to meet preclinical testing requirements, manufacturers must address any deficiencies or concerns raised by regulatory authorities. This may involve re-designing the device, conducting additional testing, or even withdrawing it from the market.

    By understanding and adhering to these regulations, manufacturers can ensure that their medical devices are safe, effective, and compliant with MDR requirements.

    DRIVING INNOVATION, DELIVERING EXCELLENCE